Technical Specification

GENERAL: 

CONTROL SOFTWARE: 

IBIX Mobile Lab for Linux – 1.0 version; multilingual user interface (available in Italian, English, French, German and Portuguese), with inter-active guide to carry out the tests, project database management, auto-matic technical report compilation, automatic updating functions via the web, calculation of estimates for lab tests, user manual 

CASE PROTECTION GRADE: waterproof 
DIMENSIONS: (55.9 x 47 x 21.6 cm) 

WEIGHT: 12.1 kg 
FUNCTIONS: 

- Optical microscopy: 
Portable USB microscope; magnification: 10X min, 150X max; built-in LED light 

- Reflectance Spectrophotometry and Colourimetry: reflectance spectrophotometer; spectral range: 410-740 nm; automatic calibration; colour space used: CIE L*a*b*; Illuminant: D65; Observer: 10°; Reference standards: EN 15886:2010 ‘Conservation of cultural property - Test methods - Colour measurement of surfaces’ 

- Measurement of water absorption under low pressure Cell to measure water absorption under low pressure by IBIX MOBILE LAB®; diameter of surface measured: 27-35-47 mm; measurement column volume: 0.2-1-2-5-10 ml; measurement on horizontal/vertical surfaces; Reference standards: NORMAL 44/93 ‘Assorbimento d'Acqua a bassa Pressione’ (Water absorption under low pressure) 

- Moisture content by gravimetric determination Weighing set; dehydration; samples processed by IBIX MOBILE LAB®; Reference standards: UNI 11085:2003 - Beni culturali - Materiali lapidei naturali ed artificiali – ‘Determinazione del contenuto d acqua: Metodo ponderale” (Cultural heritage – Natural and artificial stones - Moisture content determination. Gravimetric method) 

- Total soluble salt testing Conductivity meter and pH meter; operating range: pH from 0.00 to 14.00; EC from 0 to 3999 μS/cm; resolution: 0.01 pH; EC: 1 μS/cm; precision: ±0.05 pH; EC: ±2% FS; temperature: ±0.5 °C; automatic temperature compensation; weighing set; dehydration; samples processed by IBIX MOBILE LAB®; Reference standards: UNI 11087:2003 - Beni culturali - Materiali lapidei naturali ed artificiali - Determinazione del contenuto di sali solubili (Cultural heritage – Natural and artificial stones – Determination of soluble salt content) 

- Tests to analyse sulphate, nitrate, chloride content Photometer for transmission measurements; light source: LEDs; wave-length: 525 nm; operating range: SULPHATES: 5-150 mg/L; NITRATES: 0.1-45 mg/L; CHLORIDES: 0-210 mg/L; weighing set; dehydration; samples processed by IBIX MOBILE LAB®; Reference standards: UNI 11087:2003 - Beni culturali - Materiali lapidei naturali ed artificiali - Determinazione del contenuto di sali solubili (Cultural heritage – Natural and artificial stones – Determination of soluble salt content) 

- Ambient parameter measurement Infrared pyrometer to measure surface temperature; measurement range: -20_+270 °C; resolution: 1 °C; precision: ± 3 % of the measurement value -1 °C; measurement point (distance / size ratio): 8:1; emissivity: 0.95 

Psychrometer; measuring range: 0-100 % RH; 30...+100 °C; Resolution: 0.01 % RU; Temperature: 0.01 °C; Precision: ± 2.0 % RU at 25 °C; ± 0.5 °C at 25 °C; wet bulb temperature and dew point temperature calculation 

isuzu cikarang isuzu cikarang bekasi jakarta isuzu giga isuzu giga isuzu giga isuzu elf isuzu panther Dealer Isuzu Bekasi Cibitung Cikarang isuzu elf isuzu giga elf sparepart isuzu kredit isuzu harga isuzu promo isuzu dealer isuzu isuzu isuzu mobil isuzu bekasi isuzu online isuzu kita promo isuzu promo isuzu isuzu giga isuzu panther isuzu elf paket umroh Resep Masakan Resep Masakan